ScintCare CT 128 Intelligent Precision CT
Brief Introduction: ScintCare CT 128 is a high-end, large scale multi-slice spiral CT. Not only can meet the common demands for the basic medical institutions, but also can meet the wide demands for the large medical institutions to bear the large flux of routine outpatient, emergency, medical examination and heart scanning. ScintCare CT 128 provides the precision images due to the optimal spatial resolution.
Performance Index of the system
Detector Coverage 40mm
Detector 64 Rows
8.0MHU X-ray Tube
80KW HV Generator
128 Slices / 360°
0.39s / 360°
imA Modulation
NDI+ Iteration
Image Reconstruction Matrix: 1024 x 1024
Artifact Reduction Algorithm
Core Hardware
- Detector
Detector Type: Solid-state Ultra-High-Speed Heavy
Rare Earth Scintillator
Number of Detector Rows: 64
Number of Detectors per Row: 840
Total Number of Detectors: 53760
Minimum Physical Slices Thickness: 0.625mm
Detector Z-axis Coverage: 40mm
High-precision and High-strength Design
Modularized Digital Detector
Innovative Anti-Scatter Grid (ASG)
Digital ASIC Chip
- X-Ray Tube and Generator
Anode Heat Storage Capacity: 8.0MHU Anode Heat Dissipation: 931KHU/min
Small Focal Spot Size: ¹0.7mm²
Large Focal Spot Size: ¹1.3mm²
HV Generator
Power Rating: 80kW
mA Range: 5-667mA
kV Range: *70kV, 80kV,
100kV, 120kV, 140kV
- Gantry System
Aperture: 70cm
Gantry Tilt: ±30°(Physical)
Multipoint Temperature Monitoring
Gantry Cooling: Air
Distance from Focal Spot to ISO-Center: 558mm
Distance from Focal Spot to Detector: 948.4mm
Gantry Rotation: Belt Driven
Slip Ring Type: Low Voltage
Rotation Time: *0.37s / 360°, 0.39s / 360°, 0.5s /
360°, 0.75s / 360°, 1.0s / 360°, 1.5s / 360°,
2.0s / 360°
- Patient Table
Table pad material: Carbon Fiber, 2mm Al
Maximum Motion Range: 1800mm
Maximum Scanning Range with Extended Tabletop:
1700mm
Pitch Range: 0.25-1.75 / *0.1-2.0
Max. Speed in Horizontal Direction: 300mm/s
Motion Tolerance: 1) ±0.25mm in Z-axis Direction
2) ±0.5mm in Vertical Direction
The Min. Height of the Table: 480mm / *450mm
Max. Weight Load: 205kg / *230kg
Scan and Reconstruction
1. Data Acquisition System
Max. Number of Slices: 128 slices/ 360° Number of Detector Rows: 64 Rows Number of Detector Elements: 53760
Sampling Rate per 360°: 4096 Views / *8192 Views
Sequence Acquisition Modes: 128 x 0.625mm Spiral Acquisition Modes: 128 x 0.625mm Scan Field of View: 250mm to 500mm Reconstruction Field of View: 20mm-500mm Maximum Extended Field of Views: *700mm
Image Reconstruction and Display Matrix: 512 x 512,
*768×768, 1024 x 1024, *2048 x 2048
Image Reconstruction Speed (512×512): 60 images/s without NDI+ and 30 images/s with NDI+
Sequence Slice Thickness: 0.5mm, 0.625mm,1.25mm,
2.5mm, 5mm, 10mm
Spiral Slice Thickness: 0.5mm, 0.625mm, 1.0mm,
1.25mm, 2.0mm, 2.5mm, 3.0mm, 4.0mm, 5mm, 6.0mm,
7.0mm, 8.0mm, 9.0mm, 10mm
2. Image quality
Spatial Resolution
Technique: 120kV, 190mAs
10 lp/cm @MTF 50 %, 18mGy
13.4 lp/cm @MTF 10 %, 18mGy
20 lp/cm @MTF 0 %, 18mGy
Low Contrast Resolution
Technique: 120kV, 280mAs
3.0 mm @0.3 %, 18mGy
Technique: 120kV, 220mAs
4.0 mm @0.3 %, 14mGy
Image Noise:
Technique: 120kV, 190mAs
<=0.35% (<=18mGy)
CT Number:
Water: 0±4HU Air: -1000±10HU
CT Value: (1) Normal Range: -1024 – +3072 (2) Extended Range: -32767- +32768
3. Artifact Reduction Algorithm
Beam Hardening Compensation
Posterior Cranial Fossa Image Optimization
Streak Artifact Reduction
Motion Artifact Reduction
Helix Artifact Reduction
MAS Metal Artifact Suppression
4. Image Transfer & Networking
DICOM 3.0 Print Interface
DICOM 3.0 Output and Input Interface
DICOM Print, DICOM Store,
DICOM Query, DICOM Retrieve, Work List
HIS, RIS and PACS Interface
Raw data can be reconstructed on the workstation
DICOM Viewer on CD
5. System Software
*2048 x 2048 Imaging Reconstruction Super Matrix
Auto-voice and Integrated Intercom for
Communication between Operator and Patient
Voice Guide in 7 Languages and Recorded by User
Auto-Film
Film Printing Software
Auto-Bolus Tracking
3D-MAT (flying focal spots function)
3D Reconstruction:
MPR, CPR, MIP, MinIP, VR, SSD
Computer System
- ScintVue Console Configuration Operation System: Windows 10 in English Recon Memory: Not Less than 64GB Console Memory: Not Less than 32GB Recon Disk: 4 TB SATA + 256GB SSD Console Disk: Not Less than 2 TB
Recon Frequency: 2.4 GHz x 10 Cores x 2 CPU Console Frequency: 3.2 GHz x 6 Cores x 1 CPU LCD Display (HL2416SL): 2 x 24”, 1920 x 1200
DICOM 3.0 Network Interface
Burn Mode: CD/DVD-RW
Image Storage: about 4,000,000 Images (512 x 512)
- Workstation Configuration* Operation System: Windows 10 in English Memory: Not Less than 32GB
Hard Disk: Not Less than 2TB SATA
LCD Display (HL2416SL): 24”, 1920 x 1200
*(2 x 24”, 1920 x 1200) DICOM 3.0 Network Interface Burn Mode: CD/DVD-RW
Low Dose Technology
1. imA Modulation
With imA Modulation, CT automatically controls the output of the tube in milliamperes according
to the size of the scanner and the location of the scan. While ensuring the image quality, the patient’s radiation dose is minimized.
- Pediatric Protocol
According to the different scanning parts of
children, a special scanning protocol has been developed to avoid sensitive parts as much as possible to minimize radiation damage to children
- NDI+ Iteration
The NDI+ iterative reconstruction method directly models the noise in the data, comprehensively considers the influence of the tube ray spectrum, Bowtie material shape and detector response on
the noise, and accurately constructs the noise distribution in the CT system. In addition to the data analysis of different scanned parts, different noise reduction strategies are used in different areas. Through continuous iterative updates, image noise is reduced while maintaining a good organizational boundary.
- *Polaris Navigation
Intelligently identify human body parts, complete patient positioning including auto ISO-center determination, alignment patients with the ISO- center automatically and scout scanning according to the selected protocol.
Simplify the scanning process, realize non- contact scanning and fast workflow.
- Intelligent Puncture Localization
After scanning, when we review the images, we can choose the particular slice for puncture location and the patient table will be moved automatically.
- Smart Chest Reverse Scan
For patients who cannot hold their breath for a long time, it can significantly reduce breathing artifacts
- *CT Fluoroscopy
Equipped extra 24-inch LCD display with trolley and foot pedal in scanning room. Radiologist can control exposure and review images near the gantry.
- *Radiotherapy Planning
Equipped the extra flat tabletop and 3D laser system.
- *Myocardial Perfusion
- *CTU Analysis
- *Spine Analysis
- *Rib Analysis
- *Bone Density Analysis
- *Fat Analysis
- *4D Dynamic Analysis
- *TAVR / TAVI
- *Dual Energy Analysis